- PROBE:
- Input resistance: 2.5 GOhm / 2.5 pA
- Input bias current: ± 15 nA
- Stand-by: on/off
- ISOLATION:
- Voltage rated continuous:
- AC (50 Hz) - min. 3500 V rms
- DC - min. 4950 V rms
- Isolation mode rejection:
- AC (50 Hz) - 115 dB
- DC - 160 dB
- Barrier impedance:
- Leakage current (240 V rms, 50 Hz):
- ECG AMPLIFIER:
- Gain: 600 …7500, continuosly variable
- Input max. measurement range: 0.7 …8 mV
- Filter: band-pass, Butterworth type of the 2nd order
- frequency response (-3 dB): 0.3-100 Hz
- Notch filter: 50 Hz, Q = 6 (on/off)
- Output voltage (max): ± 10 V
- Output resistance: 1 kOhm
- TRIGGER CHANNEL:
- TTL compatible pulse (width 40 ms) triggered by the maximal positive slope of the wave.
- POWER REQUIREMENTS:
- 115 - 230 AC, 50 - 60 Hz, 25 W, fuse T 0.3A or T0.1A.
- PHYSICAL SIZE:
- 45 x 220 x 250 mm, weight cca. 1.5kg
(Specifications subject to change without notice.)
|